Flow Mesurement Knowledge

The presentation of ARTang after-sales case

Views:17     Author:Site Editor     Publish Time: 2021-08-26      Origin:Site

Customer's feedbacks about on-sit issues


The dispaly of electromagnetic flowmeter is unstable, and customers judge the flow value is  high based on past value


Troubleshooting and analysis from technical group of ARTang


Upon receiving feedback from the customer, the Anretang technical team immediately went to the customer site to examine and analyze the faults:

1. The electromagnetic flowmeter calibre is DN100, consistent with the calibre of the pipe measured in the field

2. Check the sensor coefficients of the electromagnetic flowmeter,According to the traceable source, sensor coefficients 1.7534 and the coefficient is consistent with the number calibrated by factory before shippment

3. The flow control valve is installed at the entrance of the electromagnetic flowmeter and the length of the front and rear straight pipe sections does not meet the installation requirements, resulting in a high measurement value


Solution


Modify the small signal cut point and make  the value same as the setting number before shipping.


Customer feedbacks


After modifying the small signal cut point,when no traffic running in the pipe the number of converters is normal at 0m/s



ARTANG flowmeter
Phone:
+8618868437258
Email:
luisa@artang.com
Factory:
Hangzhou City, Zhejiang
About ARTang
ARTang is the flow & level measurement manufacturer in China, specialied in electromagnetic flowmeter, liquid analysis, vortex flowmeter, level meter, widely used in water/sewage flow measurement, chemical process flow measurement, food process flow measurement, flow measurement of solid liquid, steam delivery flow measurement.